Please follow below steps to reset the chiller.
1. Restore to factory setting. 
     Turn off the chiller first. Press and keep holding ▲and▼buttons then turn on the chiller. Do not release the buttons until the controller display rE. 
 
2. To set the temperature: 
     SET" : enter / exit / confirm      
    "RST": save and exit (save the modification and exit when you have finished your modification)
    ▲or▼: to switch item / digit
 
1) Press ▲ key and hold it, then press "SET" key for 5 seconds until "0" displayed. Meanwhile, press▲key to switch to "8" and release the keys.
    After that, press "SET" key, and it displays "F0".
 
2) Press ▲ to turn to "F3" and press "SET" key, and then press ▲or ▼ to adjust "1" to "0".  Next press "SET" key to confirm and exit.
 
3) Press ▲or▼ to turn to "F0" and press ▲or ▼ to adjust "25" (supposing this is the temp. you want). Next press "SET" key to confirm and save it.
 
4) Press ▲or▼ to turn to "F9" and press ▲or ▼ to adjust "22"  or other you need (F9 value must be lower than F0, if not the chiller will stop refrigerating when the temp reached F0 value). At last,  press "RST" key to save all the modification and exit menu.